Oscar Main



 


TECHNICAL SPECIFICATION FOR OSCAR MODEL OE3000 (Oscar II)

 
Dimensions: Caseframe:

52cm (width) x 18cm (height) x 40cm (depth) (Excludes handle)

 

Weight: Caseframe: 13 kg
  Caseframe with Screen: 16.5 kg
 

Transport case:
(loaded with caseframe)

22 kg
  Handset: 0.8 kg
 

Transport case:
(loaded with handsets, probes, accessories, camera, endoscope, screen)

 

13.5 kg
Fuse Type:  

2 x T5A 250V ac

 

Supply Voltage:  

230V ac

 

Frequency of operation:

 

  28khz
Power consumption:  

450W (3 handsets) + 150W (light source)

 

Modular system:  

3 generator modules
1 (power) input module
1 optical module (endoscope light source, video camera and screen supply)

 

Insulation Classification:    
Generator: Class 1
  Handset: Type B (patient applied parts are at earth potential)
  Camera:

Type BF

 

Handset  

Stainless steel enclosure with sealed switch and cable connector suitable for autoclave sterilisation. Connecting cable provided with earthed screen.

 

Endoscope/video camera/screen  

Optional eyecup endoscopic instruments attaching to CCD camera with video image on flip-up flat TV screen on caseframe. Wipeable clear screen cover.

 

Oscar has been designed and built in accordance with BS EN 46001 Quality Assurance standard to satisfy Medical Device Directive 93/42/EEC and Part 820 of the Title 21 of the Code of Federal Regulations of the USA. CE conformance has been certified and the equipment complies with EN60601-1 and UL2601-1.

The product is covered by worldwide patents covering all international markets.





©2003 S.R.A. Developments Ltd

S.R.A. Developments Ltd
Bremridge House, Ashburton
South Devon, UK, TQ13 7JX
Telephone: +44 (0)1364 652426
Facsimile: +44 (0)1364 653589
Email: mail@sra-developments.co.uk